27
6 Calibration of the pH combination electrode
✱
Use the
and
keys to set the temperature value for the third buffer
solution and then save by pressing the
key.
Finally, the condition of the combination electrode is briefly displayed and then
the 3 point calibration is completed and the device returns to measurement
mode.
6.4.2
Calibration with any buffers
Calibration point 1
✱
Press the
key for at least 2 seconds to start the calibration.
✱
Use the
and
keys to set the value for the neutral buffer solution.
✱
Insert the combination electrode into the first (neutral) buffer solution and
gently stir.
✱
Press the
key when the measured value is stable (small triangle
appears on the display above the yellow "ok" label).
The display changes so that the temperature value for
the third buffer solution can be entered.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
NOTE!
For a sufficiently accurate reading of the slope, the pH values of the selected
buffer solutions should not be too close together (minimum gap of 2 pH from
the neutral buffer).
NOTE!
If the 2 or 3 point calibration has been selected as the calibration method,
the device expects a neutral buffer solution within the range pH 6.5 to pH 7.5
for the first calibration point.
If the 1 point calibration has been selected as the calibration method, any
buffer solution with a known pH value can be employed.
The set value for the neutral buffer solution is on the
main display of the device, "Pt.1" and a revolving
symbol appear on the subsidiary display of the
device.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...