6 Calibration of the pH combination electrode
28
✱
Use the
and
keys to set the temperature value for the first buffer
solution and then save by pressing the
key.
With 1 point calibration selected, the condition of the combination electrode
is then briefly displayed and then the 1 point calibration is completed and the
device returns to measurement mode.
✱
Use the
and
keys to set the value for the second buffer solution.
Rinsing
✱
Rinse the combination electrode and the temperature probe with de-ion-
ized (distilled) water and dry.
Calibration point 2
✱
Insert the combination electrode and the temperature probe into the sec-
ond buffer solution and gently stir.
✱
Press the
key when the measured value is stable (small triangle
appears on the display above the yellow "ok" label).
✱
Use the
and
keys to set the temperature value for the second buffer
solution and then save by pressing the
key.
With 2 point calibration selected, the condition of the combination electrode
is then briefly displayed and then the 2 point calibration is completed and the
device returns to measurement mode.
The display changes so that the temperature value
(t.1) for the first buffer solution can be entered.
The value on the main display of the device flashes,
"Pt.2" and a revolving symbol appear on the subsidi-
ary display of the device to indicate that the input of
the pH value for the second buffer solution is
expected.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
The display changes so that the temperature value
(t.2) for the second buffer solution can be entered.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...