5 Measuring
5.4.2
Determining the rH value by automatically transferring the pH value
Prerequisites
• A pH and a redox combination electrode, as well as a temperature probe if
required, must be placed in the measurement solution
• The pH combination electrode must be connected to the device
• The device must first be configured for the measurand "pH"
Measuring without temperature probe
✱
Connect the pH combination electrode to the device.
✱
Briefly press the
key.
✱
Adjust the temperature value of the measurement solution using the
and
keys.
✱
Save the temperature value with the
key.
✱
Save the stable measured pH value with the
key ("HLD" appears in the
display).
✱
Connect the redox combination electrode to the device.
✱
Configure the device for measurand "rH".
✱
Gently stir the measurement solution.
Measuring with temperature probe
✱
Connect the pH combination electrode to the device.
✱
Save the stable measured pH value with the
key ("HLD" appears in the
display).
The determined rH value appears
on the main display, the measured
value for the temperature and the
input value for the pH appear alter-
nately on the subsidiary display of
the device.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
NOTE!
Do not switch off the device until the rH measurement is finished or the
saved pH value will be deleted.
The determined rH value appears
on the main display, the input value
for the temperature and the mea-
sured value for the pH appear alter-
nately on the subsidiary display of
the device.
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
rH
mV
mV
H
pH
ATC
ok
AL Logg
pH
CAL
All manuals and user guides at all-guides.com
Содержание 202710
Страница 2: ...All manuals and user guides at all guides com ...
Страница 6: ...Contents All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 56: ...15 China RoHS 56 All manuals and user guides at all guides com a l l g u i d e s c o m ...
Страница 57: ...All manuals and user guides at all guides com ...