Remote control
68
Command
Parameter
Response of instrument
out_mode_02
2
Continual identification of controlled system whenever a new
setpoint is to be reached.
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic
out_mode_08
1
Set the control dynamics - standard
out_sp_00
xxx.x
Set working temperature.
out_sp_03
xxx.x
Set high temperature warning limit
.
out_sp_04
xxx.x
Set low temperature warning limit
.
out_sp_07
x
Set the pump pressure stage (1 ... 5).
out_hil_00
-xxx
Set the desired maximum cooling power (0 % to 100 %).
Note: Enter the value with a preceding negative sign!
out_hil_01
xxx
Set the desired maximum heating power (10 % to 100 %).
out_par_04
xxx
Band limiting during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.
out_par_09
xxx
Xp control parameter of the cascade controller.
out_par_10
xxx
Proportional portion of the cascade controller.
out_par_11
xxx
Tn control parameter of the cascade controller.
out_par_12
xxx
Tv control parameter of the cascade controller.
out _par_13
xxx
Maximum internal temperature of the cascade controller.
out _par_14
xxx
Minimum internal temperature of the cascade controller.