Preparations
28
Notice:
If an expansion vessel is not used, make sure the connector (16) is closed with a cap
nut.
If an external expansion vessel is mounted later, a part of the bath liquid should be
drained before (see chapter Emptying on page 33). It might be helpful to put the
castors at the back on a base which is approximately 10-15 mm high. So the bath
fluid can be collected in the front part of the internal reservoir.
5.5. Cooling water connection LH 50
Notice:
Cooling water circuit
Risk of oil leaking from the refrigeration system (compressor) of the recirculating
cooler into the cooling water in case of a fault in the cooling water circuit!
Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.
Even high quality heat exchangers as they are installed in our
equipment can be damaged by unsuitable cooling water.
The quality of the cooling water depends on the local conditions.
The heat exchanger may become leaky due to corrosion or it may
become clogged due to particulate matter
Caution:
Securely attach all tubing to prevent slipping.
IN
OUT
12 mm inner dia.
Example:
PVC-tubing
•
Supply cooling water via the inlet (18).
•
Conduct the cooling water via the outlet (17) in the
respective sink or return flow circuit.
Cooling water connectors
G3/4" external thread
Tubing
12 mm inner dia. tubing
Cooling water see page 12