67
Command
Parameter
Response of instrument
in_mode_05
none
Temperature system in Stop/Start condition:
0 = Stop
1 = Start
in_mode_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
in_par_01
none
Time constant of the external bath.
in_par_02
none Internal
slope.
in_par_03
none
Time constant of the internal bath.
in_par_04
none
Band limiting (max. difference between the temperatures in the
internal bath and external system).
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_par_09
none
Xp control parameter of the cascade controller.
in_par_10
none
Proportional portion of the cascade controller.
in_par_11
none
Tn control parameter of the cascade controller.
in_par_12
none
Tv control parameter of the cascade controller.
in_par_13
none
Adjusted maximum internal temperature of the cascade
controller.
in_par_14
none
Adjusted minimum internal temperature of the cascade controller.
out commands: Setting parameters or temperature values.
Command
Parameter
Response of instrument
out_mode_01
0
Use working temperature
out_mode_02
0
No identification.
Temperature control by using the stored parameters.
out_mode_02
1
Single identification of controlled system after the next start.
Содержание Presto Plus LH 47
Страница 20: ...Operating controls and functional elements 20 Rear...
Страница 41: ...41...