51
Examples:
Profile No 1:
Step
Setpoint
Time
Edit
Compile profiles:
•
A flashing segment indicates that a number needs to be
entered.
Under submenu "Edit Profile" enter a profile number.
Six profiles may be stored (nos. 0 to 5).
•
Then programme the desired values for each section.
Use the keypad to set section number, target temperature and
time period. Set each entry with enter
.
When the program is running, only sections having complete
information for target temperature and time period are
considered.
It makes sense, to leave out section numbers in the profile, in
order to use them later for corrections in the profile.
Important:
If a time of 00:00 is set for a profile, the profile is continued
with the next section only after the setpoint temperature
(±0.2 °C) is reached.
•
Press
escape
to return to the previous menu level.
Example:
Step
(No.)
1 3 5 6 9 11
14
Setpoint (°C)
50 50 100
100
80 80 20
Time
(h.m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30
Step
Edit Profile: 1
>Step: 3
Setp: 50:00°C
Time[h.m] 00:02
Display sections:
1. Use the cursor keys
to select the submenu
"Step", enter the desired number and press enter
.
or
2. If the submenu „Set“ is selected, section by section can be
shown in ascending order by the P-key
.
3. Each previously set value is displayed.
Содержание Presto Plus LH 47
Страница 20: ...Operating controls and functional elements 20 Rear...
Страница 41: ...41...