Menu functions
44
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing
the start of the instrument directly by pressing the mains power
switch or using a timer.
Possible parameters:
on - AUTOSTART on
off - AUTOSTART off
Warning
:
For supervised or unsupervised operation with the AUTOSTART function, avoid
any hazardous situation to persons or property.
The instrument does no longer conform to N.A.M.U.R. recommendations.
Take care you fully observe the safety and warning functions of the instrument.
If the setpoint is given via the serial interface, no AUTOSTART is possible.
STAND-BY
>Stand-by input
External stand-by for emergency switch-off
(connector - see page 6)
Possible parameters:
no - stand-by input is ignored
yes - stand-by input is active
> Language
There are two options for the language of the DIALOG-DISPLAY
(LCD): German or English.
Possible parameters:
German (deutsch)
English (engl).
S xx.xx
ID 0
←
>IdentNo
The identification number identifies the removable control
module RD.
Assign identification numbers if more than one Presto
®
temperature system is available.
Adjustable parameters:
0 ... 999
Indicated also on the VFD-Info-Display.