E n v i r o n m e n t a l c o n d i t i o n s a c c o r d i n g t o I E C 6 1 0 1 0 - 1 :
•
Use only indoor.
•
Altitude up to 2000 m - normal zero.
•
Ambient temperature: see Technical specifications
•
Air
humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
•
Max. mains fluctuations of ±10 % are permissible.
•
Overvoltage category II
•
Pollution degree 2
Caution:
The unit is not suitable for use in explosive atmosphere
Protection class according to IEC 60 529: IP31
The unit corresponds to Class I
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).