Device modes
System response
in_mode_01
Return value 0. The value 0 stands for the control temperature
setpoint as only one temperature setpoint can be set (downward
compatibility for older devices)
in_mode_02
Selftune:
0 = Selftune disabled
2 = Selftune Enabled
in_mode_03
Operating mode setting for the EPROG input:
0 = 0…10 V
1 = 4…20 mA
in_mode_04
Operating mode setting for the temperature control:
0 = Internal
1 = External
in_mode_05
Operating mode set for the temperature control system:
0 = Stop
1 = Start
in_mode_08
Controller dynamics set with internal temperature control:
0 = Aperiodic
1 = Standard
in_mode_11
Actuating variable source:
0 = Circulator
1 = Serial
2 = Analog (EPROG)
IN PAR
System response
in_par_00
Set sensor difference between working temperature sensor and
safety temperature sensor
in_par_01
Set time constant of the external bath TE
in_par_02
Set internal gradient value SI
in_par_03
Set time constant of the internal bath TI
in_par_04
Set optimization parameter CoSpeed
in_par_05
Ration of cooling/heating capacity pk to ph
in_par_06
Set control parameter Xp of the internal controller
in_par_07
Set control parameter Tn of the internal controller
in_par_08
Set control parameter Tv of the internal controller
in_par_09
Set control parameter Xp of the cascade-controller
in_par_10
Set control parameter Xpu of the subordinate controller (cascade
temperature control)
in_par_11
Set control parameter Tn of the cascade-controller