![Julabo Magio MX Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737047.webp)
Initial operation
47
7.6 Fill device
This section describes how the device should be filled with bath fluid during
initial operation.
Specifications for filling volume can be found in the technical data.
The drain valve is closed.
The device is switched off.
1. Remove the bath lid.
2. Half fill the bath with bath fluid.
The bath fluid expands with increasing temperature and can overflow.
With decreasing temperature, the low liquid level protection can be
triggered and interrupt the temperature control process.
3. Switch the unit on and start the temperature control application.
4. Watch the level indicator and, if necessary, adjust the level by refilling or
draining.
Once the working temperature has been reached and the sample inserted,
the level of bath fluid in the bath tank should cover the heating coil of the
heating circulator or the cooling coil of the refrigeration unit.
5. Close the bath opening with the bath lid.
The device is filled with bath fluid.
Содержание Magio MX
Страница 1: ...Bridge mounted bath and heating circulators Original operating manual 1 950 1750 us V03 11 2021...
Страница 100: ......