![Julabo Magio MX Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737053.webp)
Operation
53
8.3.5 Set physical units
The physical units for temperature, pressure and flow can be set in the menu.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Settings>
.
3. Set the desired units for temperature, pressure and flow.
The physical units are set.
8.4 Unit account
In the menu option
<Unit account>
the access rights for user groups can be
managed.
Basic users and advanced users can log in with this menu option. Logging in
activates the respectively assigned rights.
The administrator manages the access rights of the two user groups and can
change passwords.
8.4.1 Change password
The password ensures that only authorized persons have access to defined
device functions. The administrator password is required to change a password.
Passwords can be created and changed for the following user groups:
• Basic users, factory setting password: 0000
• Extended users, factory setting password: 00000
• Administrators, factory setting password: 000000
The device is switched on.
1. Call up the
<Main menu>
.
2. In the submenu
<Unit account>
call up the menu option
<Settings>
.
An input field will open.
3. Enter the administrator password and confirm your input.
Factory setting for the administrator password: 000000.
4. Select one of the three user groups.
5. Press the
[Change password]
button.
An input field will open. The new password may have up to 19 digits.