Operation
61
8.6.3 Adjust controller
In the menu <
Adjust controller>
the individual control parameters for
internal and external control can be set.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Determine thermodynamics>
and the menu
option
<Adjust controller>
.
3. Activate internal or external control.
This results in the respective adjustable control parameters. External control
is only possible with an externally connected Pt100 temperature sensor.
4. Switch to the submenu
<Control performance>
.
5. Set the desired parameters.
6. For internal control: Activate the aperiodic temperature sequence or the
standard temperature sequence.
The control parameters are set and immediately active.
8.6.4 Limit cooling capacity and heating capacity
In the menu
<Limits>
the cooling capacity and heating capacity are limited.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Determine thermodynamics>
and the menu
option
<Adjust limits>
.
3. Set the limit values for the cooling capacity and the heating capacity.
The values are set and are immediately active.
8.6.5 Adjust pump
The performance of the pump can be set in one-percent increments.
The device is switched on.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Determine thermodynamics>
and the menu
option
<Adjust pump>
.
3. Enter the desired value for pump capacity in percent.
The minimum value is defined internally. The maximum value is limited to
the adjustable maximum pump capacity in the menu
<Adjust safety>
.
4. Confirm your input.
The set value is immediately active. The pump is set.