11.1.1 IN commands
IN commands retrieve device parameters.
Process values
System response
in_pv_00
Actual value
in_pv_01
Current variable (%)
In_pv_02
Current temperature of the external Pt100 temperature sensor
in_pv_03
Current temperature of the temperature safety sensor
in_pv_04
Current setting of the high temperature safety function
in_pv_06
Current pressure. EPROG input must be set to pressure
in_pv_07
Current flow. EPROG input must be set to flow
in_pv_16
Current bath fluid filling level (%)
Setpoints and warning
limits
System response
in_sp_00
Set working temperature
in_sp_03
Set high temperature warning limit
in_sp_04
Set low temperature warning limit
in_sp_05
Working temperature set via EPROG
in_sp_06
Setpoint temperature set via watchdog function
in_sp_07
Set pump stage (1 … 4)
in_sp_10
Set actuating variable specification via serial interface
in_sp_11
Set temperature unit
0 = °C
1 = °F
in_sp_12
Set pressure unit:
0 = bar
1 = °F
in_sp_13
Set flow unit:
0 = l/min
1 = gpm
in_sp_27
Set pump setting
in_sp_28
High temperature alarm limit
in_sp_29
Low temperature alarm limit