![Julabo Magio MX Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/julabo/magio-mx/magio-mx_original-operating-manual_2030737064.webp)
Operation
64
7. Start the terminal program on the computer.
8. Enter the interface parameters into the terminal program.
9. Use the terminal program to select the COM port of the circulators or the
connected adapter, and establish a connection.
Remote control via the RS485 serial interface is activated. You can now
remote control the circulators with interface commands via the terminal
program. See appendix for interface commands.
8.7.5 Set EPROG input
The EPROG input can be used to send an external voltage or current signal to
the circulator. Depending on the setting, the circulator interprets the input
signals as a setpoint temperature, power capacity, flow or pressure, and can
output the values as a display on the start screen.
Please note the technical data of the REG/EPROG socket.
The device is switched on.
The remote control mode is deactivated.
An external signal transmitter is connected via the REG/EPROG socket.
1. Call up the
<Main menu>
.
2. Call up the submenu
<Connect unit>
then menu option
<Interfaces>
.
3. From the submenu
<Analog module>
select the menu option
<EPROG>
.
4. Select the input variable
[Setpoint]
,
[Performance]
,
[Flow]
or
[Pressure]
.
5. Select the signal type voltage or current.
6. Use the external signal transmitter to go to the lower voltage or current
limit.
The value is displayed in the menu as the current value (V or A).
7. Tap the
[Low value]
button and enter the low input value (setpoint,
performance, flow or pressure).
The value is displayed in the menu as the current value (°C, °F, %, l/min,
gpm, bar, psi).
8. Repeat steps 6 and 7 for the upper limit value.
The EPROG input is set. In the submenu
<Settings>
, the values for display
on the start screen can be selected. The setpoint temperature or
performance can also be used as an external setpoint value or as an
actuating variable.
Содержание Magio MX
Страница 1: ...Bridge mounted bath and heating circulators Original operating manual 1 950 1750 us V03 11 2021...
Страница 100: ......