User instructions
Torque ratchets
FB-EV 065 GA Drehmomentratsche Rev. 3 (Version: 26.10.2020) |
(Translation: 26.10.2020)
Page 1 of 5
Thank you for choosing to buy one of our dental products.
Carefully read through these instructions before each use and store them somewhere that is easy for
the user and for the relevant qualified personnel to access.
Carefully read the warning information indicated by this warning symbol. Improper use of the products can lead to
serious injury to the patient, user or any third parties.
To ensure that the condition and functionality is faultless for the intended use, the specifications of the following instructions
must be complied with. Please be aware that handling implements improperly can negatively affect their service life and safe-
ty.
The medical devices are not delivered in a sterile state and must be prepared and sterilised by the user
according to the following instructions before the first and before any subsequent use.
1 Delivered condition, storage of brand-new products
As a general rule, the devices should be stored in a dry place and protected against dust, chemical fumes or components.
The products are delivered without tension at approx. 10 Ncm. This must also be maintained as the general storage condition to ensure
optimum functionality and a longer service life.
2 Product details
This user manual is only applicable for the products specified below.
Product description / product group:
Torque ratchet
Catalogue and order numbers
1000201 1000206 1000210 1000216 1000500 1000700 TWR
301312
1000202 1000207 1000211 1000217 1000501 1000701 P48935 518-1000203-F
1000203 1000208 1000214 1000218
1000702 CD
518-1000701-F
1000205 1000209 1000215
4572R
Each individual part only belongs to the delivered implement. Exchanging components is not permissible (even with identical
implements) and requires a new inspection to be performed by the manufacturer before the torque function can be used.
This product is a medical device and is only intended for use by trained dental specialists.
The relevant employee must be sufficiently qualified in accordance with statutory regulations, and with the training and hygiene require-
ments, for the repreparation of the device.
It’s the user’s responsibility to select suitable procedures and employees relating to the product.
2.1
Intended use
This torque ratchet is to be used for the temporary insertion and removal of screws and for the insertion of implants, as well as for loosening
them in defined torque ranges for dental applications in the fields of implantology, osteosynthesis, surgery and prosthetics. The torque
function can also be “blocked”. In the blocked position, higher torques can be used for insertion, as well as removal.
For implements with ranges up to 80 Ncm → Using with loads above 100 Ncm can damage the implement.
For implements with ranges up to 100 Ncm → Using with loads above 120 Ncm can damage the implement.
2.2
Contraindication
Special contraindications can only be seen in connection with operation procedures. Therefore, the user is responsible for the selection of
suitable methods and settings in accordance with the individual anatomical characteristics of their patients. The torque ratchet must not be
used in case of any intolerance or allergies against conventional surgical stainless steels.
2.3
Combination with tools or other products
There are adaptors available which allow you to use this torque ratchet with many different tools. Adaptors manufactured by Josef Ganter
GmbH are generally suitable. The user must ensure that they choose the suitable size for the intended tool connection.
Due to the large number of possible combinations (including combinations with the end tools of other manufacturers), there are always
detailed technical specifications available under
in the download area of the current catalogues.
When using adaptors produced by other manufacturers, their guidelines regarding the compatibility of said adaptors with these
user instructions, at least with regards to the connection size to be used, the intended user and the repreparation, is to be
checked. We are not liable for any damage caused by combinations with third-party products, unless the problem concerns a
manufacturer that was expressly named in one of the catalogues mentioned in this paragraph.
3 Use / handling
Immediately before each use, the product must be checked for any possible signs of wear, loss or limitation of function or
corrosion. In addition, the implement must be assembled correctly.
Damaged products or products with any of the aforementioned faults must be immediately scrapped and must not be used in
this condition!
If the sterile packaging of products (after being prepared by the user) is damaged, the products should not be used and must undergo
another repreparation according to these instructions.
3.1
Possible default settings
Prosthodontic setting – torque function: The desired torque range can be continuously set via the spring using the adjusting nut. The setting
can be seen on the scale of the scale sleeve.
Surgery setting – blocked function: Turn the adjusting nut to the scale mark (infinity symbol). Do not tighten excessively.