Battery Power Line II Instructions for Use
DePuy Synthes
5
Notes:
–
Always wear personal protective equipment
(PPE) including safety goggles when handling
with the BPL II system.
–
To avoid injuries, the locking mechanism of the
tool has to be activated before every manipula-
tion and when laying the tool down, i.e. the mode
switch has to be in the off position.
–
Only place the tool in an upright position when
changing attachments or cutting tools intraopera-
tively. The handpiece must be laid on its side
when not in use in order to avoid the risk of be-
ing dropped or contaminating other instruments.
–
If the machine has been dropped it should be
checked carefully for damage. In the event that
any damage is visible, do not use it anymore and
send it to the DePuy Synthes service center.
–
The tool may only be operated with a fully
charged battery. Therefore, ensure that the bat-
tery is charged in good time. We recommend
that the battery is placed into the charger imme-
diately after surgery.
–
The batteries must never be sterilized, washed,
rinsed or dropped. This would destroy the battery
with possible secondary damage (explosion
hazard!). Only use original Synthes batteries.
Further information can be found on page 18ff.
–
Never place the BPL II in a magnetic environ-
ment since the machine might start unintention-
ally.