Introduction
Universal Battery Charger II
Battery Power Line II Instructions for Use
DePuy Synthes
13
The Universal Battery Charger II (05.001.204) in-
cludes four independent charging bays. Each charg-
ing bay has three slots; the Battery Power Line II
battery (530.630) fits into the top slot.
Note:
In order that the BPL II battery can be recognized
and charged by the UBC II, firmware version 14.0* is
required. If necessary, send the charger to a DePuy
Synthes representative for a firmware update.
*Sticker as shown on the underside of the charger
with firmware version 14.0:
For further information on the Universal Battery
Charger II please consult the relevant Instructions for
Use (036.000.500) or contact your local DePuy Syn-
thes office.
The BPL II battery cannot be charged with the Uni-
versal Battery Charger (item number 530.600 or
530.601).
1
Charging bays (4)
2
Symbols for battery type
3
ON/OFF display
4
Control display for each charging bay
5
Ventilation holes
6
Ventilation holes
7
Power switch
8
Fuses: 2 x 5 AT / 250 V
9
Power cord connection
10
Symbol for
BPL I and II batteries (530.620 or
530.630)
10
SW-Rev. 14.0
2013/01/09