Care and Maintenance
Manual Cleaning Instructions
Battery Power Line II Instructions for Use
DePuy Synthes
41
1. Remove debris
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free
cloth or soft-bristled brush to assist in removing
gross soil (Fig. 1). For cannulations of the handpiece
and attachments, the cleaning brush (516.101)
should be used.
Notes:
–
Do not use pointed objects for cleaning.
–
Important: Never immerse the handpiece, batter-
ies, battery casing or attachments in aqueous
solutions or in an ultrasonic bath or use pressur-
ized water as this will cause damage to the sys-
tem.
–
Brushes and other cleaning tools shall be either
single-use items or, if reusable, be decontami-
nated at least daily using a solution as detailed
below in section “3. Spray and wipe”.
Brushes shall be inspected before daily use and
discarded if they have degraded to the point
where they may scratch instrument surfaces or
be ineffective due to worn or missing bristles.
2. Manipulate moving parts
Manipulate all moving parts such as triggers, sliding
sleeves, attachment release rings, saw blade cou-
pling and switches under running tap water to loosen
and remove gross debris.
3. Spray and wipe
Spray and wipe the device using a neutral pH enzy-
matic solution for a minimum of 2 minutes (Fig. 2).
Follow the enzymatic detergent manufacturer’s direc-
tions for correct temperature, water quality (i.e. pH,
hardness) and concentration / dilution.
Figure 1
Figure 2
Содержание BPL II
Страница 2: ......
Страница 4: ...2 DePuy Synthes Battery Power Line II Instructions for Use...
Страница 75: ......