Operating Instructions
Battery Pack (battery casing with inserted battery)
18
DePuy Synthes
Battery Power Line II Instructions for Use
Charging, storing and using batteries
Charging
Only use the Synthes Universal Battery Charger II
(05.001.204) to charge the battery.
Using a charger
that does not originate from Synthes can damage the
battery.
In order that the BPL II battery can be recognized
and charged by the UBC II, firmware version 14.0 is
required. If necessary, send the charger to a DePuy
Synthes representative for a firmware update.
The BPL II battery cannot be charged with the Uni-
versal Battery Charger (item number 530.600 or
530.601).
The batteries should always be charged before use.
Place the battery into the charger immediately after
surgery.
Charge the batteries within an ambient temperature
range of 10°C / 50°F to max. 40°C / 104°F.
Keep the charger and the batteries clean and in a
cool and dry place.
Detailed information on the Universal Battery
Charger II can be found in the Instructions for Use
(036.000.500).
Storage
Always recharge the Li-Ion Battery for Battery Power
Line II (530.630) after each use. Do not store an
empty battery as this will shorten the life span and
will not be covered by warranty.
When the battery is not in use, store it in the Synthes
Universal Battery Charger II (05.001.204). This will
guarantee that the battery is always fully charged
and ready to use.
The Universal Battery Charger II should always be
turned on when a battery is in the charging bay.
This ensures availability of charged batteries.
Do not expose batteries to heat or fire. Avoid storage
in direct sunlight.
Keep the charger and the batteries clean and in a
cool and dry place.
Use
Do not remove battery from its original packaging
until required for use.
Do not drop or apply force to the battery.
This will destroy it with possible secondary damage.
Do not use any battery which is not designed for use
with the equipment.
Do not use a faulty or damaged battery, as this may
damage the power tool.
Do not short-circuit a battery.
Do not store or transport batteries haphazardly in a
box or drawer where they may short-circuit each
other or be short-circuited by other metal objects.
This can damage the batteries and generate heat,
which can cause burns.
Batteries give their best performance when they are
operated at normal room temperature (20°C / 68°F
+/- 5°C / 9°F).
The power tool may only be operated with a fully
charged battery. Therefore the batteries should al-
ways be charged before use.
Place the battery into the charger immediately after
surgery.
Only insert the battery pack directly before using the
power tool. This saves battery energy and prevents
having to change it during surgery.
Follow the information in the “Care and Mainte-
nance” section starting on page 39 as well as the
Instructions for Use of the Synthes Universal Battery
Charger II (036.000.500).