Chapter 8
TROUBLESHOOTING
Problem
Possible Cause
Solution
1. Chain is too
loose/tight
Incorrect number or
combination of links
for proper scanner
configuration.
Refer to the
ROTIX
setup chart
(see Two Probe
Chain Configuration Setup Chart on page 52)
or
(see Four Probe Chain Configuration Setup
Chart on page 53)
for required number of
links for the diameter of pipe/tube that is to be
scanned. Ensure the correct outer diameter
measurement of the pipe/tube. Reset the
scanner with the correct number of links.
Buckle is incorrectly
setup.
Adjust tightness of buckle
(see Setup of ROTIX on
a Scanning Surface on page 34)
2. Insufficient
probe contact.
Scanner not set
properly.
Reconfigure the scanner as per instructions
(see
Probe Holder Adjustment on page 16)
8.1. Technical Support
For technical support contact Jireh Industries
(see “Jireh Industries Ltd.” on page 1)