4.15.5. Probe Holder Left/Right Conversion
To reverse the probe holder, follow these steps:
NOTE:
To perform this operation the 1.5 mm hex wrench (Fig. 29) is required.
1.
Ensure the probe holder is in latched, upper position
(Fig. 95)
.
2.
Using the supplied 3 mm hex driver
(Fig. 27)
, unscrew the yoke pivot screw
and remove yoke
(Fig. 105)
.
3.
Loosen the probe holder arm adjustment knob and the arm clamp screw.
Slide the probe holder arms off the yoke
(Fig. 106)
.
4.
Flip the yoke 180° and swap the probe holder arms
(Fig. 107)
.
5.
Place the pivot buttons on the inside of the probe holder arms
(Fig. 108)
using a 3/8 in wrench
(Fig. 28)
.
Fig. 105
- Unscrew yoke pivot screw
Fig. 106
- Remove probe holder arms
Fig. 107
- Flip yoke and reverse arms
Fig. 108
- Attach arms & move buttons