Jenway 6300 Ser Man
10
2.0
About ‘Quick Reference’
This section contains a selection of the key information that is
often forgotten or difficult to find when required. Use Quick
Reference as a memory jogger, but for more information check out
the references to the main sections on each point.
2.1
Specification
Also see Section 1.2 of the Instruction Manual.
Wavelength Range
320nm to 1000nm
Wavelength Resolution 1nm
Wavelength Accuracy
+/-2nm
Spectral Bandwidth
8.0nm
Light Source
Tungsten Halogen
Optics
Single Beam
Transmittance
Range; 0 to 199.9%
Resolution; 0.1%
Absorbance
Range; -0.300A to 1.999A
Resolution; 0.001A
Concentration
Range; -300 to 1999
Resolution; 0.1, 1
Units; ppm, mgl
-1
, gl
-1
, M, %, blank.
Factor
0 to 999.9 / 1000 to 9999
Photometric Accuracy
+/-1%
Photometric Noise
Less than 1%
Stray Light
Less than 0.5%T
Photometric Stability
Better than 1% per Hour (after warm up)
Readout
Custom LCD
Outputs
Analogue (0 to 1999mV) & RS232 Serial
Supply Voltages
115/230 V a.c.
Power
Less than 50W
Dimensions
365 (w) x 272 (d) x 160 (h) mm
Weight
6Kg