6-5
(2)
Time program functions and their number
1) Wavelength 1:WAVELENGTH
2) CD output range 2:CD RANGE
3) UV output range 3:AU RANGE
4) Spectrum measurement 4:MEASURE
5) Auto zero 5:AUTO ZERO
(3)
Time program input method:
The input sequence is as follows: time, function, value relative to the function.
1) From the initial conditions screen (Step 0 screen), press [
▼
] key to
advance to the Step 1 screen.
2) Press [EDIT/ENTER] (time blinking).
3) Enter the time in minutes (minimum increment: 0.1 min).
[xx][EDIT/ENTER]
The function will be blinking.
4) Select a function by using numeric key or [
▼
]. Input the function by
pressing [EDIT/ENTER]. The function will appear and the value will
blink, awaiting input.
5) Input the desired value by using numeric key or [
▲
] and press
[EDIT/ENTER] to enter the value.
6) Press [
▼
] to move to the next program step. Repeat this procedure
until all steps are entered. After the final step completed, press
[MONIT] to go to the program mode monitor screen. ( see Fig. 6.4)
Note: Input steps need not be entered in the proper time sequence. When
the [MONIT] is pressed in order to return to the monitor screen, steps
that are out of time sequence will be rearranged into the proper
sequence.