1.2 Specifications
Light source :
Hg-Xe lamp(150W)
Measurment wavelength range : 220-420 nm
Measurement range:
1000 mdeg (300nm) in 12 steps
Spectral bandwidth :
20 nm
Wavelength accuracy :
±5 nm
Sensitivity (S/N=2) :
0.05 mdeg/cm (290 m),
0.1 mdeg/cm (240 nm)
0.1 g (Camphor-10-Sulfonic acid, 290 nm)
0.01 g (Pantoy Lactone, 240 nm)
Baseline drift :
0.1 mdeg/cm/h (at 290nm)
Gain:
x1,
x10
Response :
FAST, STD, SLOW
Flow cell path length :
25 mm(1,5,10 mm optional)
Signal outputs :
Simultaneous outputs of CD
(circular dichroism) and UV absorption
CD, UV recorder output = 10mV full scale
CD integrator output = 1V/deg
UV integrator output = 1V/AU and 1V/2AU
Spectral measurement :
Simultaneous measurement of CD and UV
spectra
Spectral memory :
220~420
nm(both
UV
and
CD),
Memory:10files
Power requirement :
100-240 VAC, 400VA
Dimeniton:
390(W)x 450(D)x250(H)mm
Weight :
approx. 21kg