AP-1600T-USB | AP-3200T-USB | User Manual (Ver. 2.0)
Main Functions
- 51 -
Manual Configuration
1. Select the index in
BlemishCompensationIndex
.
You can select from 1 to 200. However, configure the indexes in order starting with the smallest index. If
you skip indexes while configuring settings, interpolation may not be performed.
2. Specify the pixel points for interpolation using the
BlemishCompensationPositionX
and
BlemishCompensationPositionY
settings. Each point is saved to the camera's internal memory as you
configure them.
You can configure values that are within the total effective pixel area. Specify pixels for which
interpolation is not necessary as -1. If 0 is specified, the first line or first pixel will be interpolated.
Note:
By executing
BlemishCompensationDataClear
(specify sensor)
BlemishCompensationIndex
, you can return a specific pixel correction setting to the default value
(storage not required).
3. Execute
BlemishStore
. Blemish compensation data will be stored.
4. Set
BlemishEnable
to
True
, and execute interpolation. If
False
is specified, defective pixel interpolation
is disabled.