PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
78
Export binary phase data
This command is supported only when viewing the selected plot’s phase record
contents using
Measurement
→
Phase difference (
p
)
or
Measurement
→
Frequency-
difference (
f
)
. It will save a binary file containing phase difference data in seconds,
represented as a block of double-precision IEEE 754 values in little-endian (Intel) format.
If a selected or zoomed area has been defined by dragging with the mouse, only that
portion of the phase record will be exported.
As with
File
→
Export ASCII phase data (
x
)
, the default file suffix is .CSV. This is usually
inappropriate for binary data; you should specify a suffix appropriate to your
application. Exported binary phase data may not be re-imported into TimeLab.
Export phase data to Stable32 (
Ctrl-x
)
This command is useful for transferring data from the selected plot’s phase record
directly into a third-party analysis application such as
. Stable32 is a popular application for offline data analysis. It supports many
statistical features and display options beyond the limited set implemented by TimeLab.
This command works by exporting the selected plot’s phase record to a temporary file
using the same functionality as the
File
→
Export ASCII phase data (
x
)
command. If a
selected or zoomed area has been defined by dragging with the mouse, only that
portion of the phase record will be exported.
When the command is issued, a dialog box allows you to specify the location of the
Stable32 executable or other third-party program, as well as the command line to be
passed to it. The command line may contain various predefined variables described in
the dialog box’s help text, including placeholders for the name of the temporary file and
the data rate (tau zero) in samples per second. Upon pressing the
Launch Stable32
button, the temporary file will be generated and the specified program launched.
TimeLab will then continue to run normally.
Temporary files created by TimeLab are cleaned up when TimeLab exits, so it’s advisable
to leave TimeLab running until after the third-party analysis application has exited.
Содержание PHASE STATION 53100A
Страница 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Страница 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Страница 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Страница 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Страница 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Страница 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Страница 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Страница 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Страница 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Страница 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Страница 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Страница 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Страница 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Страница 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Страница 143: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 143 Figure 66 5 MHz residual phase noise and AM noise...
Страница 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Страница 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...