/rate:<1, 10, 100, 1000>
By default,
STREAM.EXE
configures the instrument driver to make 10 measurements per
second. The baseband measurement bandwidth is half of this interval, or 5 Hz by
default. The 53100A drivers limits the minimum bandwidth to 5 Hz, even when /rate:1
is used.
You can use the /rate option to select 1 reading per second to minimize network
bandwidth and/or disk space for use with extremely stable sources. Conversely,
measurements of drift-prone sources may require /rate:100 or /rate:1000 to avoid
acquisition failures.
Note that TimePod measurements made at 1 reading per second (0.5 Hz BW) are likely
to fail if insufficient warmup time is allowed for the DUT, the reference source, or the
instrument itself.
/timestamp:<s, MJD>
/sep:<character>
These related options cause
STREAM.EXE
to timestamp each phase-difference or
frequency reading sent to files or network clients.
By default, no timestamps are emitted. When /timestamp:MJD is specified, the
Modified Julian Date (MJD) associated with each phase-difference or frequency value
will be written as the first numeric field on each line, followed by a space character
(ASCII 32), then the measurement value itself. MJD timestamps are generated with six
digits of decimal precision, resulting in roughly 10 unique MJD values per second.
You can also use /timestamp:s to emit a simple numeric timestamp at one-microsecond
precision. These timestamps begin at 0.000000 for the first reading acquired.
In either case, the difference between successive timestamps will always be equal to the
reciprocal of the /rate parameter, whose default is 10 readings per second.
The /sep option may be used to specify an additional separator character that will
appear before the space between the timestamp and measurement value. If the client
requires comma-separated values, for example, /sep:, would be useful. Any numeric
value used as the /sep: argument is treated as an ASCII code; e.g. /sep:9 generates a tab
character.
Example:
stream timepod.tll /rate:1 /format:F /timestamp:MJD /sep:,
Содержание PHASE STATION 53100A
Страница 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Страница 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Страница 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Страница 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Страница 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Страница 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Страница 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Страница 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Страница 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Страница 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Страница 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Страница 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Страница 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Страница 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Страница 143: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 143 Figure 66 5 MHz residual phase noise and AM noise...
Страница 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Страница 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...