The phase difference chart’s overall slope is a representation of the fractional difference
between the nominal and actual frequencies. In the example above, as in the earlier
show freqdiff
example, the reported DUT input frequency is 10.1 MHz while the actual
frequency from the test source is just under 10.123456 MHz. The readings above
advance by about 2.322 milliseconds per second, corresponding to the fractional
frequency differences returned by the
show freqdiff
command. Continuing the
example, if the nominal input frequency had been specified with
set inputfreq
10.123456
,
/inputfreq:10.123456
, or
/roundfreq:0.000001
, the slope of the phase
differences would be almost zero.
See the
measurelinear
and
removelinear
commands for information on calculating and
removing the linear trend from the phase-difference chart.
-
show phaserate
Returns the phase data rate set by the
/phaserate
command-line parameter or the most
recent
set phaserate
command. See the description of the
/phaserate
command-line
parameter for more information.
-
show referencefreq
Returns the reference frequency used for frequency counter and phase/frequency
difference calculations. See the description of the
/referencefreq
command-line
parameter for more information.
-
show spectrum
Returns a list of offset frequencies and dBc/Hz values corresponding to the phase noise
spectrum in the measurement in progress (Figure 87):
Following the PSD data, the current instrument noise floor estimate will be returned
(Figure 88):
Figure 87
Содержание PHASE STATION 53100A
Страница 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Страница 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Страница 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Страница 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Страница 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Страница 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Страница 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Страница 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Страница 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Страница 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Страница 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Страница 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Страница 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Страница 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Страница 143: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 143 Figure 66 5 MHz residual phase noise and AM noise...
Страница 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Страница 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...