PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
127
Acquisition options (
Ctrl-F7
)
This command brings up a dialog box that configures TimeLab’s autosave and deferred-
acquisition functionality.
When deferred-acquisition mode has been toggled on by the
Acquire→
Enable deferred
acquisition (
Ctrl-d
)
command, data from subsequent measurements will be discarded
until the measurement is “triggered” manually with
Trigger deferred acquisition(s)
(
Enter
)
or by cancelling deferred execution with a second
Enable deferred acquisition
(
Ctrl-d
)
command. The
Acquisition delay
field in the dialog box above provides an
additional way to trigger a deferred measurement. By setting a nonzero
Acquisition
delay
value in seconds, a deferred measurement will trigger itself automatically once a
specified number of seconds has elapsed after initiation. With the default delay value of
zero seconds, manual triggering is necessary in order to begin collecting data. Refer to
the
Enable deferred acquisition (
Ctrl-d
)
command description for more information.
The autosave feature helps guard against data loss due to forced operating system
updates, power failures, or other unexpected events that may occur during long-term
measurements. When the
Autosave interval
field in the dialog box has been set to a
value greater than or equal to 1 minute, the .TIM files associated with all subsequent
measurements can be periodically saved to disk at the specified interval, just as if the
user had issued a
File→Save image or .TIM file
(
s
)
command manually.
The autosave interval can be specified at any time before or after the beginning of
acquisition. However, only .TIM files can be automatically saved during the
measurement, not image files. Additionally, each .TIM file to be automatically saved
must be saved manually at least once in order to assign it a valid filename.
Figure 59
Содержание PHASE STATION 53100A
Страница 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Страница 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Страница 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Страница 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Страница 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Страница 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Страница 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Страница 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Страница 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Страница 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Страница 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Страница 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Страница 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Страница 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Страница 143: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 143 Figure 66 5 MHz residual phase noise and AM noise...
Страница 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Страница 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...