150
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
Figure 10-13. Typical QTH lamp.
6.
Carefully align the two legs of the lamp into the two holes in the socket.
There is no polarity on this lamp.
7.
Gently push the bulb down into the socket until it is fully seated.
8.
Replace the cover and tighten the SHC screws to lock the cover closed.
9.
Reset the hour meter by pressing the reset button located on the front of
the M-2000 Detector and Lamp Power Supply Module.
Note:
When a new lamp is installed, it must be aligned so that the
maximum amount of light is available for sample measurement.
The optimal position should not change much from lamp to lamp,
but fine adjustment is needed to achieve best results.
Lamp Alignment
1.
Turn on the lamp power supply. Allow the lamp to warm up for at least
20 minutes.
2.
Ensure the Base is in the 90° (straight-through) position and the Z-
position is raised such that the sample stage will not block the beam from
hitting the receiver unit’s aperture.
3.
Select
Hardware
>
Controls
>
General
> ‘Align Sample’. Choose No when
prompted for the full system alignment.
4.
Locate the Average Signal reading (bottom of the alignment screen) and
make a mental note of the signal value.
5.
There is a small adjustment screw located on the bottom of the lamp
housing. This screw controls the vertical position of the lamp inside the
lamp housing. Using the 1/16” Allen wrench, adjust the screw until the
signal reading is maximized.
Содержание M-2000
Страница 1: ...M 2000 Spectroscopic Ellipsometer Hardware Manual Auto Angle ESM 300 Base CompleteEASE Software...
Страница 2: ...2 2023 J A Woollam Co M 2000 Hardware Manual...
Страница 4: ......
Страница 9: ...M 2000 Hardware Manual 2023 J A Woollam Co ix...
Страница 10: ......
Страница 50: ......
Страница 152: ...152 2023 J A Woollam Co M 2000 Hardware Manual...
Страница 172: ......
Страница 175: ...M 2000 Hardware Manual 2023 J A Woollam Co 175 M 2000V I M 2000X I...
Страница 177: ...M 2000 Hardware Manual 2023 J A Woollam Co 177...