146
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
Note:
Turning off the lights in the room may help facilitate the next few
steps.
4.
While holding a business card in the path of the beam, adjust the
adjustment screws using a 5/64” English Allen (2mm Metric) wrench until
the beam is visible (a properly aligned lamp produces a very bright,
bluish-white beam).
Note:
Only slight adjustment is needed to give the beam a uniform spot on
the business card. It may be necessary to iterate between the two
adjustment screws for best uniformity and intensity.
5.
Once the beam is visually bright, use the ellipsometer to perform the final
alignment of the lamp: Open the alignment screen by selecting,
Hardware
>
Controls
>
General
>‘Align Sample’. Choose No when
prompted for the full system alignment. If your system has Auto Tip-Tilt,
ensure 'Receiver’ is selected in the
Detector
panel at the top of the
screen.
Figure 10-9. Alignment screen.
6.
Adjust the lamp adjustment screws to maximize the Average Signal
Intensity displayed at the bottom of the Alignment Screen
7.
Adjust the Source unit tilt stage slightly so the beam’s spot is centered on
the receiver unit’s aperture.
8.
Using the Receiver unit tilt stage, adjust so the cross hair is centered on
the alignment screen.
Содержание M-2000
Страница 1: ...M 2000 Spectroscopic Ellipsometer Hardware Manual Auto Angle ESM 300 Base CompleteEASE Software...
Страница 2: ...2 2023 J A Woollam Co M 2000 Hardware Manual...
Страница 4: ......
Страница 9: ...M 2000 Hardware Manual 2023 J A Woollam Co ix...
Страница 10: ......
Страница 50: ......
Страница 152: ...152 2023 J A Woollam Co M 2000 Hardware Manual...
Страница 172: ......
Страница 175: ...M 2000 Hardware Manual 2023 J A Woollam Co 175 M 2000V I M 2000X I...
Страница 177: ...M 2000 Hardware Manual 2023 J A Woollam Co 177...