i-ALERT2 Application Guide
27 of 64
MOUNTING METHODS
There are three methods that can be used to mount the i-ALERT2 device
to a machine. They are shown in the figure below.
On new pumps and pumps repaired through ITT PRO Services, a slot will be milled into an appropriate location on the
bearing housing and the device will be secured with a screw. Alternatively, a hole can be drilled and tapped and a small
adapter plate can then be used to secure the device. Lastly, the device can be epoxied directly the machine via the
mounting plate. Make sure to use a good quality, rigid epoxy. Do not use RTV or Silicon to mount the device.
CAUTION:
Always wear protective gloves. The equipment and the i-ALERT2
device can be hot.
Mill a Slot
Drill and Tap
Epoxy
Screw: ¼-28 x 1.125in
Screw: ¼-28 x 1.5in
Screw: ¼-28 x 1.125in
Slot: 2.25in L x 1.6in W
Tap: ¼-28 UNF x ¼in deep
Epoxy: application specific
COMMISSIONING
STEPS TO ACTIVATE THE I-ALERT2 EQUIPMENT HEALTH MONITOR:
1. Remove the sticker.
2. The i-ALERT2 device
will look for light. Note: If in a dark environment use light source to activate.
3. When activated a sequence of flashing LEDs will start to indicate that the unit is powered on.
4. When sequence is completed the green LED will flash every 5 sec under normal operating conditions
Figure 23: Mounting Methods
Содержание i-ALERT2
Страница 1: ...Application Guide...
Страница 2: ...i ALERT2 Application Guide 2 of 64...
Страница 32: ...i ALERT2 Application Guide 32 of 64 Figure 27 Devices in range and out of range...
Страница 38: ...i ALERT2 Application Guide 38 of 64 Figure 33 Export trend data by printing or attaching to email...
Страница 58: ...i ALERT2 Application Guide 58 of 64 Figure 57 Route Report Trend Details Screen...
Страница 63: ...i ALERT2 Application Guide 63 of 64...
Страница 64: ...i ALERT2 Application Guide 64 of 64 G iALERT2AppGuide en US 2016 10...