Recommended Separation Distances
The EndoPAT
TM
x is intended for use in an electromagnetic environment in which radiated
radiofrequency disturbances are controlled.
The user and/or installer of the unit can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile radiofrequency communications equipment
(emitters) and the EndoPAT
TM
x, according to the maximum output power of the equipment, as
recommended in the table below.
Table 6 - from IEC 60601-1-2:2014
Recommended separation distances between
portable and mobile RF communications equipment and the EndoPAT
TM
x (EPX)
Rated maximum output power of
transmitter
Watts [W]
Separation distance according to frequency of transmitter (in meters)
Meters [m]
150kHz to 80MHz
d
= 1.17√P
80MHz to 800MHz
d
= 1.17√P
800MHz to 2.5GHz
d=
2.3√P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.73
1
1.17
1.17
2.3
10
3.7
3.7
7.3
100
11.7
11.7
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.
Table 6a
Recommended separation distances between
Power bus or an appliance wire and the EndoPAT
TM
x (EPX)
Rated maximum current (I) in a
power bus or an appliance wire
Amperes [A]
Separation distance (in meters)
Meters [m]
R
= I/2H=I/18.8
0.1
0.005
1
0.05
10
0.53
100
5.3
1000
53.2
NOTE 1 The power frequency magnetic field immunity of H=3 A/m is met.