Itamar Medical Ltd.
Endo
PAT
TM
x Device
78
Operation Manual
Test
frequency
(MHz)
Band a)
(MHz)
Service a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
Immunity
Test level
(V/m)
Compliance
level
(V/m)
930
CDMA 850,
LTE Band 5
1720
1 700 –
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
3,
4, 25; UMTS
Pulse
modulationb)
217 Hz
2
0.3
28
28
1845
1970
2450
2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulationb)
217 Hz
2
0.3
28
28
5240
5 100 –
5 800
WLAN 802.11
a/n
Pulse
modulationb)
217 Hz
0.2
0.3
9
9
5500
5785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME
SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it
would be worst case.