Table 4 - from IEC 60601-1-2:2014
Recommended separation distances between portable and mobile RF communications equipment and the EndoPATx
The
EPX
is intended for use in the electromagnetic environment specified below; The customer or the user of the
EPX
should assure that it is
used in such an environment.
Rated maximum
output power of
transmitter [
W ]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
outside ISM bands
150 kHz to 80 MHz
in ISM bands
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0.01
0.12
0.2
0.4
1
0.1
0.37
0.64
1.3
2.6
1
1.17
2
4
8
10
3.7
6.4
13
26
100
11.7
20
40
80
Table 5 - from IEC 60601-1-2:2014
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The
EPX
is intended for use in the electromagnetic environment specified below; The customer or the user of the
EPX
should assure that it is
used in such an environment.
Test
frequency
(MHz)
Band a)
(MHz)
Service a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
Immunity
Test level
(V/m)
Compliance
level
(V/m)
385
380 –390
TETRA 400
Pulse
modulationb)
18 Hz
1.8
0.3
27
27
450
430 – 470
GMRS 460,
FRS 460
FM c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
28
710
704 – 787
LTE Band 13,
17
Pulse
modulationb)
217 Hz
0.2
0.3
9
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
Pulse
modulationb)
18 Hz
2
0.3
28
28
870