The product is designed to comply with EN 61010-1 and can be flash
tested. It is fitted with radio frequency interference suppressors.
Therefore it is recommended that only a D.C. test be performed.
Performing flash tests repeatedly can damage insulation
This equipment contains no user-serviceable parts. Refer all repairs to
qualified service personnel. Contact Isothermal Technology or one of our
appointed distributors for details of approved service outlets
1.3 Powering Up Your microK
The microK operates on any standard AC electrical supply (88-264V RMS at 47-
63Hz) so unless your supply is unusual you can simply use the power cord
provided to connect your microK to a suitable electrical outlet.
The power switch is located at the rear of the instrument, immediately below the
IEC connector. When you turn your microK on it will go through a standard
Windows CE™ boot sequence and then display a Window containing the
Resume
button. Pressing
Resume
restarts the microK in the configuration it was
in prior to the last power down, you will then see the Main Window and be ready
to operate your instrument.
Before pressing the
Resume
button, make sure that any thermometers attached
will not be damaged by the sense currents that will be applied (the last values
used). If there is a problem, disconnect your sensors before proceeding and then
reconnect them after changing the sense current to the required values.
1.4 A Quick Tour of Your microK
On the front panel of your microK you will find the input terminals for the three
measurement channels, the touch screen/colour display and a USB connector: