is one that your have created in the Thermometer Database this box will be fixed
and will be the conversion type defined for that thermometer.
Range box:
Use the up/down buttons to select the required range. The microK
has three ranges for each thermometer type. Use the 0.125V range for all
thermocouples as this provides the most suitable input voltage range. The
resistance ranges offered depend on the sense current (use the ellipses button by
the
Current
box in this tab to change the current). Choose a resistance range that
is greater than or equal to both the highest thermometer resistance and reference
resistor value you intend to use.
Reference Channel box (PRTs and thermistors only):
Use the up/down
buttons to select the reference resistor used in the resistance measurement; either
one of the internal reference resistors or the channel to which an external
resistance standard is connected. Alternatively press the ellipses button to open
the Reference Channel window and select the channel directly. The
External
button in this window allows you to select one of the microK’s three external
channels or any expansion channel provided by a microsKanner.
Reference Resistor box (PRTs and thermistors only):
If an internal reference
resistor was selected in the
Reference Channel
box, this box will show the
calibrated value of that resistor. If one of the input channels was selected in the
Reference Channel
box, this box will contain a list of the reference resistors in
the microK’s database (entered using the
Edit Resistors
button in the
Instrument
tab – see section 3.7.2) and “Uncalibrated”. Select the required resistor using the
up/down buttons. Select “Uncalibrated” if you only want to make a resistance
ratio measurement between two channels. The ellipses button opens a list of all
available reference resistors and is useful if you have a large number of entries in
the database.
Reference Junction box (thermocouples only):
This contains
0°C, 0.01°C and
a list of any input channels that have been configured to measure temperature.
Choose 0°C or 0.01°C if you are using an external reference junction in an ice-
point bath or water triple point respectively. Choose one of the measurement
channels if you want to measure the temperature of the reference junction with a