Keyboard window and enter a serial number. The
Shift
key toggles the alpha
keys between upper and lower case.
Calibration Due box:
This box contains the date when calibration is due for the
selected reference resistor. If you are creating a new reference resistor entry, the
ellipses button by the box becomes active. Press this button to open a Date Entry
window and enter a date. The ellipses buttons by the
Day
,
Month
and
Year
boxes
in the Date Entry window open a Numeric Keypad window. Dates are checked
for validity (including leap years) for dates up to 31
st
December 2099.
Value box:
This box contains the calibrated value for the selected reference
resistor. If you are creating a new reference resistor entry, press the ellipses
button to open a Numeric Keypad window and enter the calibrated resistance
value (in ohms).
New button:
Press this button to make a new thermometer entry into the
microK’s database (ellipses and up/down buttons by the boxes then become
active).
Delete button:
Press this button to delete the selected reference resistor (or the
current entry for a new reference resistor) from the database.
OK button:
Press
this button to
confirm the changes and close the Window.
3.7.3 Calibration
Press the
Calibration
button to open a Date Entry window and set a recalibration
date for your microK. The microK will warn you if you attempt to use it past this
recalibration date (you can disregard the warning and continue to use it, but the
warning will continue to re-appear every hour until the out-of-calibration issue is
resolved).
3.7.4 Restart Software
Press the
Restart Software
button to restart the microK software. This is useful if
you have added or removed microsKanners and want to re-invoke the automatic
discovery process without cycling the power.