type in the command
CAL:CURR
current
, where
current
is the measured value.
Now type in
TEST:CURR 1
again to update the current source. The ammeter
should now read 1mA ± 0.0003mA.
Check the other current ranges by typing in the following commands and
checking that the current is within the given limits:
command
current limits
TEST:CURR 0.1
0.09975 to 0.10025 mA
TEST:CURR 10
9.975 to 10.025 mA
TEST:CURR -0.1
-0.09975 to -0.10025 mA
TEST:CURR -1
-0.9975 to -1.0025 mA
TEST:CURR -10
-9.975 to -10.025 mA
7.2.2 Calibrating the 400
Ω
Internal Reference Resistor
Connect a stable external resistance standard with a nominal value of 100
Ω
to
channel 1. Ideally, this should be a Wilkins resistor with a recent (traceable)
calibration certificate maintained at a constant temperature in an oil bath.
Configure the microK to measure a Default PRT at 1mA against the internal
400
Ω
reference on the 500
Ω
range with channel 1 (other channels disabled). Set
the units to ratio (Rt/Rs) with 1 sample per reading and 100 readings in the
rolling statistics (see section 3.6). Select the
Single
tab and clear the statistics.
After the microK has collected at least 100 readings (shown
as Mean of last…
in
the
Single
tab, record the
Mean
value and check that the standard deviation
(
S.D.
) is less that 0.000000100. Calculate the value of the internal resistor from:
ratio
reference
exernal
R
=
400
Adjust the value assigned to the internal 400
Ω
reference resistors by typing in
the command:
CAL:REF205
R400
The value entered can be read back using
CAL:REF205?
.Check the new value
assigned to the internal 400
Ω
reference resistor (see section 7.1.2.2).