![Isotech micro-800 Скачать руководство пользователя страница 77](http://html.mh-extra.com/html/isotech/micro-800/micro-800_user-manual_2093605077.webp)
©
Page 77 of 97
Issue: 1.04
10.2.1
SCPI Numeric Suffices
In order to support multiple input channels, commands can include numeric
suffices. These are represented by a hash (#) in the command notation, for
example:
MEASure:VOLTage:<channel#>?
In addition to the 3 input channels on the front panel (which are assigned the
suffices 1, 2 and 3) the internal reference resistors are specified using numeric
suffices as follows:
channel#
Internal reference resistor
201
1
Ω
202
10
Ω
203
25
Ω
204
100
Ω
205
400
Ω
In order to measure the voltage on channel 3, the command used would be:
MEAS:VOLT3?
10.2.2
Parameters
If a command requires parameter(s), these follow the command and are separated
from it by a space. If more than one parameter is required, each parameter is
separated by a comma (,). For example, the command to measure a four-terminal
resistance on channel 1 against a reference resistor on channel 2 using the 100
ohm range and a sense current of 1mA would be:
MEAS:FRES1:REF2? 100,1
10.2.3
Units
Numeric parameters may optionally include units. The standard multiplying
prefixes may also be used to indicate decimal multiples or fractions. For
example, the following are all equivalent.
CURRENT 0.001A
CURRENT 1ma