Page 67 of 97
Issue: 1.04
calibration certificate. Repeat the procedure for calibrating the 400
Ω
internal
reference (see section 7.2.2), but configure the microK to measure a Default PRT
at 10mA against the internal 1
Ω
reference on the 13
Ω
range on channel 1. The
command to adjust the value assigned to the 1
Ω
internal reference is:
CAL:REF201
R1
Where
R1
is the value calculated for the 1
Ω
internal reference.
7.2.7 Calibrating Voltage Zeros
Short out the 4 measurement terminals on all 3 channels using a piece of un-
insulated copper wire (use pure copper wire to avoid generating unwanted
thermal EMFs) for each channel:
V
I
Type in the following commands to remove the zero/offset correction for all
three channels:
CAL:OFFS1 0
CAL:OFFS2 0
CAL:OFFS3 0
Configure the microK to measure a Default Thermocouple on the 0.125V range
with units of Volts on all 3 channels. Use 1 sample per reading and 100 readings
in the rolling statistics (see section 3.6). Select the
Multi
tab
and clear the
statistics for all 3 channels. After the microK has collected at least 100 readings
(shown as
Mean of last…
in the
Multi
tab, record the
Mean
for each channel and
check that the standard deviation (
S.D.
) is less that 0.000000050 (0.05µV).
Now adjust the zero offsets using the commands: