5.3
Display settings
Customer specific settings for data to be displayed can be achieved by using
the appropriate menu items.
5.3.1
Main PV
The main Process Value (PV) is the primary measurement data.
5.4
Output settings
5.4.1
Serial interface RS 232
The RS 232 serial interface can be used to transmit data on-line or to down-
load the integral datalogger content. The settings can be found in the
Serial
Communication
submenu.
5.5
download software
Software can be provided for downloading the contents of the datalogger and
communication with the flowmeter.
5.6
Wall thickness measurement (WTM)
Optional sensor probes to measure pipe wall thickness are available. The
IFX-P200 will recognise a connected probe when entering the Setup or WTG
Wizards, the measurement mode or the Scope function. Use the Setup Wiz-
ard or Installation menu to set the pipe material. Select “Start Measurement”.
The IFX-P200 will recognise the probe and display the measurement screen.
Wall thickness will be shown when the sensor is in good acoustic contact
with the pipe.
5.6.1
Wall Thickness Gauge (WTG) Wizard
To confirm pipe thickness and sound speed, select the "WTG Wizard" from
the "Quick Start" menu. Enter the approximate expected thickness as "Ref-
erence THK" and select "Calibrate".
The screen displays the received acoustic pulse and values for the signal
strength, the transit time, the reference thickness, the expected sound speed,
the measured thickness at the reference sound speed and the measured
sound speed at the reference thickness (top to bottom).