28
Passes
Select from list ↑↓
Auto
1...16
WTG Wizard
Reference THK
Calibrate
Stored Setup
Load
Load from list
Save
Save to list – use alphanumeric keys to enter
name
Delete
Delete from list
Start
Measurement
Sensor type
Indication of sensor type and serial number if au-
tomatically detected, otherwise
select from list ↑↓
As Setup Wizard
Sensor fre-
quency
SP1, only for special, unrecognised sensors
Wedge angle
SP2, only for special, unrecognised sensors
Wedge c-
speed 1
SP3, only for special, unrecognised sensors
Wedge c-
speed 2
SP4, only for special, unrecognised sensors
Crystal offset
SP5, only for special, unrecognised sensors
Spacing offset
SP6, only for special, unrecognised sensors
Zero flow offset
SP7, only for special, unrecognised sensors
Upstream off-
set
SP8, only for special, unrecognised sensors
K factor
Calibration factor (slope)
Sensor
placement
Adjust sensor position
Installation
Pipe
Material
Select from pipe material list ↑↓
Outside diame-
ter
10 ... 3000 mm
Wall thickness 0.5 ... 75 mm
Inner diameter 10…3000 mm
C-speed
600 ... 6554 m/s (transverse sound speed)
L-Speed
600 ... 8000 m/s (longitudinal sound speed)
Circumference 31.4 ... 9420 mm
Roughness
0.0 ... 10 mm
Medium
Fluid
Select from fluid list ↑↓
Kinematic
viscosity
0 ... 30000 mm
2
/s
Dynamic
viscosity
0 ... 30000 mPa/s
Density
100 ... 2000 kg/m
3
C-speed
800 ... 3500 m/s
Temperature
-30 ... 300 °C