Installation location
Select an installation location following the recommendations in Table 1 and
try to avoid measuring:
●
in the vicinity of deformations and defects of the pipe,
●
near welding seams,
●
where deposits could be building up in the pipe.
For a horizontal pipe:
Select a location where the transducers can be mounted on the side of the pipe, so that the sound waves emitted by the
transducers propagate horizontally in the pipe. In this way, the solid particles deposited on the bottom of the pipe and
the gas pockets developing at the top will not influence the propagation of the signal.
Correct
Incorrect
For a free inlet or outlet pipe section:
Select the measuring point at a location where the pipe cannot run empty.
Correct
Disadvantageous
Correct
Disadvantageous
For a vertical pipe:
Select the measuring point at a location where the liquid flows upward to ensure that the pipe is completely filled.
Correct
Incorrect
Look for a sensor installation location with sufficient straight pipe to obtain
accurate measurements. Please refer to Table 2 as a guideline for recom-
mended distances from disturbance sources.
Table 1: Recommendations for sensor mounting location