Manual No. 0158A
24/68
IWB
−
IWX Series (Overseas Use) Service Manual
Measurement Conditions Setup Table
Address Weight
Item
Data
1
0: 0
1:0.0
2
Decimal point
position
2: 0.00
3:0.000
4
Minimum indication
0: 1 (At dual / Multi Interval 1
−
2 switching)
4: 2 (At dual / Multi Interval 2
−
5
−
10 switching)
110A
8
8: 5 (At dual / Multi Interval 5
−
10
−
20 switching)
C: Not available
110B
-
Accuracy (resolution) 0: 1/500
1:1/600
2:1/750
3:1/1000
4: 1/1200 5:1/1500 6:1/2000
7:1/2500
8: 1/3000 9:1/4000 A:1/5000
B:1/6000
C: 1/7500 D:1/10000 E:1/12000
F:1/15000
1
2
4
Switching method
0: Single Range LED
Stable indication only (IWX)
1: 1/3000 Multi Interval (Stable and Upper/Lower limit
indication with IWB)
2: Single Range
LED Upper/Lower limit
indication only (IWX)
3: 1/3000 Multi Interval (Stable and Upper/Lower limit
indication with IWB)
4: Single Range (bathroom scale specifications)
5: 1/3000 Dual Range (A/B range switching)
6: Single Range (lb/kg switching)
7: 1/3000 Multi Interval (lb/kg switching)
111A
8
1
Weight unit
(applicable only to
transmission
function)
<111A> <111B>
0
+
0 : kg (kg-lb)
8
+
0 : g
0
+
1 : lb
8
+
1 : oz
2
4
111B
8
Filter setup
Do not change the
data, for these data
determine the
stability (flickering).
(Cut off)
(Notch)
(Output rate)
0: 0.66 Hz
2.50 Hz
400 ms
2: 0.84 Hz
3.20 Hz
312 ms
4: 1.05 Hz
4.00 Hz
250 ms
6: 1.31 Hz
5.00 Hz
200 ms
8: 1.68 Hz
6.40 Hz
156 ms
A: 2.10 Hz
8.00 Hz
125 ms
C: 2.62 Hz
10.00 Hz
100 ms
E: 3.35 Hz
12.80 Hz
78 ms