IGB/IGX Series (Overseas Specifications) Service Manual
2.9 C8 Mode: E2ROM Data Reading
Data from address 000 to address 127 can be read.
Operation Display
1.
Starting Test Mode
Press
ON/OFF
.
Press and hold
ZERO(+)
while all segments are lit.
2.
Entering C8 Mode
Press
ZERO(+)
seven times.
Or, press
TARE(-)
twice.
3.
Press SET.
To change an address
Press
ZERO(+)
to increase an address.
Press
TARE(-)
to decrease an address.
Address 000
Address 110
Address 127
TARE key
C8 Starting address
ZERO key
Address 3 digits
−
Data 1 byte
4.
Press COLUMN to return to Test Mode.
E2ROM Data
Address Item Data
Range
Remarks
Data set in Program
Mode of Operation
Manual.
10 byte × 10 data
0 to 99
Preset data (10 PLUs)
Upper limit
17bit
Lower limit
17bit
Tare amount (A range)
18bit
Tare amount (B range)/Single weight 20bit
Check sum
8bit
0 to 60000
Reference
Reference
Refer to IGB/IGX
Operation Manual.
100A
Scale mode data Capacity (bit1)
lb mode (bit2)
Counting (bit3)
0 to F
100B Reserved
-