Manual No. 085-3435-07
21/45
IGB/IGX Series (Overseas Specifications) Service Manual
2.8 C7 Mode: Weighing Condition Setting
Note
Setting Country No. and Scale No. determine the weighing conditions for securing certified
specifications of the country.
Individual contents can be changed in this mode.
Specifications must be selected when used overseas.
Operation Display
1. Starting Test Mode
Press
ON/OFF
.
Press and hold
ZERO(+)
while all segments are lit.
2. Entering C7 Mode
Press
ZERO(+)
six times.
Or, press
TARE(-)
three times.
3. Press SET.
To change data:
Select the desired data by pressing
ZERO(+)
to increase,
or
TARE(-)
to decrease the address.
To fix the data or advance the address:
Press
SET
.
3-digit a A (Upper data) - Data
3-digit a B (Lower data) - Data
4. After changing data
Push the Memory Switch (SW1) located on the main
board.
(This does not work when there was no data change)
5. Press
COLUMN twice to return to Test Mode.
Reference
Reference
Addresses for weighing condition settings are from 110 to 115.
Refer to the next page.