Manual No. 085-3435-07
14/45
IGB/IGX Series (Overseas Specifications) Service Manual
2.2.3 Operation Procedure
Operation Display
1. Starting Test Mode (C1)
Press
ON/OFF
.
Press and hold
ZERO(+)
while all segments are lit.
Press
SET
.
2. Setting Country Number
To increase the number, press
ZERO(+)
.
To decrease the number, press
TARE(-)
.
Press
SET
.
[Example] Asia = 3
3. Setting Scale Number
When setting “60kg”, select “08”.
Press
SET
.
4. Selecting Weight at Span Adjustment
The following weight can be selected:
kg:
1/1.0
→
1/2.0
→
1/2.5
→
1/5.0 of weighing capacity
lb:
1/1.0
→
1/2.0
→
1/2.5
→
1/5.0 of weighing capacity
・
Pressing
ZERO(+)
selects the next item in the above
sequence, and pressing
TARE(-)
selects the previous item.
kg
unit: Same weight is used as weighing
capacity.
[Example]
Selecting “kg” unit and “1/2.0” for weighing
capacity “150kg” will require the total 75kg
weight.
kg
unit: Half weight is used of weighing
capacity.
To change between “lb” and “kg”, press
ZERO(+)
and
TARE(-)
at the same time.
lb
unit: Same weight is used as weighing
capacity.
5. Zero Point Adjustment
After selecting the weight amount, ensure that nothing is
placed on the weighing platter, then press
SET
.
The previously set zero point A/D data will be displayed.
Zero point indicator
W
will light up.
Press
ZERO(+)
to set “20000” counts forcibly.
Zero point indicator
W
will go off.