Ishida Setup & Calibration Manual

 

1.1

NOVA-II 

 

Follow the steps below to set up and calibrate the NOVA-II scale.

 

IMPORTANT!

 

  All work performed on the NOVA-II, including setup and calibration, must be done by an authorized

scale distributor to ensure compliance with NTEP requirements.

 

Figure 1-1. Bottom view of NOVA-II scale

 

Step

Action

Display

1

Turn the scale off.

2

Turn the scale on and immediately press the 1 key.

3

Press the 1 key then the TARE key.

C     0

4

Press TARE after verifying the country code: 20 = United States (15 lb, 30 lb, and 60 lb)
To change the country code, enter 20 20, then press TARE.

C1-1   20   0

5

Press TARE after verifying the Capacity Code:

• 2 = 30 x 0.01 lb
• 9 = 15 x 0.005 lb
• 7 = 60 x 0.02 lb

 

NOTE:

 

 

 

XXXX

 

 = capacity, z = code. To change, enter the code, then press TARE.

C1-2   

 

XX.XX

 

    

 

z

 

6

With no weight on the scale, press ZERO to set 2000 counts. 

 

NOTE:

 

 

 

nnnn

 

 = any number.

10  

 

nnnnn  nnnn

 

7

Place the full capacity weight on the scale as determined in Step 5.

10   2000   0

8

Press TARE. 

 

NOTE:

 

 

 

nnnn

 

 = any number

10   

 

nnnnn   nnnn

 

9

The center display should read 32000 ± 1. If not, repeat the above starting with Step 6.

10  32000  

30000

10

Remove the weights and the weigh platter.

10  32000  

30000

11

Tip the scale on edge and remove the brass screw. See the diagram below. 

 

NOTE:

 

 

 

nnn

 

 = 

any number.

10   

 

nnnnn   nnnn

 

12

Use a 

 

non-metallic

 

 device to press the memory switch on the circuit board. See the 

diagram below.
The scale beeps and temporarily displays “ok.”

10   

 

nnnnn nnnn

 

10    

 

s

 

    ok

13

Replace the brass screw and the weigh platter. Turn off the scale.

10    2000    0

14

Turn the scale on again and test the calibration by placing a known weight on the platter.

0.00   0.00   0.00

Memory Switch

Содержание BC-3000

Страница 1: ...51860 Ishida Industrial Non System Retail Products Setup Calibration Manual ...

Страница 2: ......

Страница 3: ...out notice November 2002 Contents Introduction 1 1 0 Retail Scales 1 1 1 NOVA II 2 1 2 BC 3000 3 2 0 Industrial Scales 5 2 1 MTX MTB 6 2 2 MTE Series 8 2 3 MG 3000 9 2 4 CWS Series 11 2 5 CWS Series 12 2 6 NCS Series 13 2 7 QBX Series 14 3 0 Balances 15 3 1 QB Series 16 3 2 MB Series 17 3 3 MB RE Series 18 3 4 IB Series 19 3 5 SB Series 20 ...

Страница 4: ......

Страница 5: ...cians responsible for installing and servicing these products Authorized distributors and their employees can view or download this manual from the Ishida distributor site at www ishidaretail com 1 0 Retail Scales Rice Lake Weighing Systems currently sells and distributes the following models of Ishida retail scales Nova II BC 3000 The following pages describe the setup and calibration procedures ...

Страница 6: ...z code To change enter the code then press TARE C1 2 XX XX z 6 With no weight on the scale press ZERO to set 2000 counts NOTE nnnn any number 10 nnnnn nnnn 7 Place the full capacity weight on the scale as determined in Step 5 10 2000 0 8 Press TARE NOTE nnnn any number 10 nnnnn nnnn 9 The center display should read 32000 1 If not repeat the above starting with Step 6 10 32000 30000 10 Remove the w...

Страница 7: ...point 0 2000 5 Initial A D value should read 2000 1 If No Press ZERO key again If Yes Proceed to step 6 0 2000 6 Place 30 lb weight on platter 30000 32000 7 Is A D change value within 30000 2 If NO Press the TARE key If YES Proceed to step 8 Remove weight from platter and return to step 4 30000 32000 8 Remove weight and weigh platter from unit to perform data registration 9 Remove white plastic sc...

Страница 8: ...4 Ishida Setup Calibration Manual ...

Страница 9: ...Weighing Systems sells and distributes the following models of Ishida industrial scales MTX Series MTB Series MTE Series MG 3000 Series CWS Series NCS Series QBX Series The following pages describe the setup and calibration procedures for these products ...

Страница 10: ... step To Review 1 After the operator enters C8 see table above the input displays C4 C1 setting 2 If the C4 C1 settings are not correct turn the scale off then on again and re enter the settings If the C4 C1 settings are correct press the COUNT key 3 The unit displays the C8 C5 settings 4 If the C8 C5 settings are not correct turn the scale off then on again and re enter the settings If the C8 C5 ...

Страница 11: ... 1 0 30 lb 15 kg 150 lb 75 kg G 2 0 24 lb 12 kg 120 lb 60 kg G 2 5 DO NOT USE L 1 0 THROUGH L 5 0 THE SCALE WILL NOT CALIBRATE NOTE Use a test weight of 100 capacity for greatest accuracy Press the COUNT key to go to the next step The initial A D value displays nnnn 5 Press the ZERO key twice to obtain the value 2000 2000 6 Add weights to equal the value selected in Step 4 After the scale stabiliz...

Страница 12: ... initial A D value NOTE A normal A D initial value is between 6000 and 10000 count 8280 5 Press the ZERO key to set the A D zero point value 2000 6 Place a 300 Lb weight on the scale platter then press the TARE key to set the A D full capacity value 70040 2 7 Remove the weight then press the ZERO key to set the A D zero point value 2000 8 Place the weight on the platter again NOTE If the displayed...

Страница 13: ...to change the digit position Press the CHG VAL key to increment the value Cb xx 7 Press the SET key Cb 20 8 Press the CHG VAL key to select the capacity 11 15 x 0 005 Lb 10 6 x 0 002 Lb CAP xx 9 Press the SET key CAP 11 10 With no weight on the scale press the ZERO key until the display shows 2000 counts xxxx 11 Place the full capacity weights on the scale as determined in Step 9 above 2000 12 Pre...

Страница 14: ... Press the TARE key Cb 20 9 Press the CHG VAL key to select the capacity b 15 x 0 005 Lb A 6 x 0 002 Lb CAP x 10 Press the TARE key CAP b 11 With no weight on the scale press the ZERO key until the display shows 2000 counts xxxx 12 Place the full capacity weights on the scale as determined in Step 9 2000 13 Press the TARE key once xxxxx 14 The span counts should read 17000 1 If not repeat the proc...

Страница 15: ...off then on again and re enter the settings 2 B N2 N1 14 The X s represent the internal reference counts Press the ZERO key xxxx 15 With no weight on the scale press ZERO until the display shows 2000 counts 2000 16 Place the full capacity weights on the scale as determined in Step 5 above 2000 17 Press the TARE key yyyy 18 The system performs calibration CAL 19 The span count should read 7000 If n...

Страница 16: ... correct press TARE If not turn the scale off then on again and re enter the settings 2 B N2 N1 14 The X s represent the internal reference counts Press the ZERO key xxxx 15 With no weight on the scale press ZERO until the display shows 2000 counts 2000 16 Place the calibration weights on the scale as follows 5 Lb capacity 5 Lb weight 10 Lb capacity 10 Lb weight 19 Lb capacity 25 Lb weight 2000 17...

Страница 17: ...n the scale press ZERO to set 2000 counts NOTE nnnn any number 10 nnnnn nnnn 7 Place the full capacity weight on the scale as determined in Step 5 10 2000 0 8 Press TARE NOTE nnnn any number 10 nnnnn nnnn 9 The center display should read 32000 1 If not repeat the above starting with Step 6 10 32000 30000 10 Remove the weights and weight platter 10 32000 30000 11 Tip the scale on edge and remove th...

Страница 18: ...then release the key 2 Press and hold the T key press the MODE key then release both keys at the same time 3 un it appears Press the T key to increase the value of the right digit in this order 1 2 5 Select the desired until system number see table below 4 Press the MODE key to call up on 0 The QB automatically performs zero adjustment Do not leave any weight on the pan 5 After the QBX performs ze...

Страница 19: ...s Rice Lake Weighing Systems sells and distributes the following models of Ishida precision balances QB Series MB Series IB Series SB Series The following pages describe the setup and calibration procedures for these products ...

Страница 20: ... digit in the following order 0 1 2 9 A 6 d repeat from 0 Select the desired unit system number see table below 4 Press the F key to call up on 0 The QB automatically performs a zero adjustment Do not leave any weight on the pan 5 After the QB performs zero adjustment on FS automatically appears Gently place the designated calibration weight on the center weigh pan see table below for designated c...

Страница 21: ...50 of the standard If 2 Err appears on the display the display tolerance 0 4 is exceeded or the scale may be defective Contact your Ishida representative Step Action 1 Press and hold the F key until Func CAL appear in sequence then release the key 2 Press the ZERO key While holding it down press the F key then release both keys The display blinks on 0 to indicate zero adjustment is being performed...

Страница 22: ...pears on the display after returning the knob to WEIGH this indicates that the error at span varies too much from that of the factory set value Figure 3 1 Calibration Knob Step Action 1 Press the S key and release SA CAL appears in the display 2 The display blinks CAL 0 which indicates zero adjustment is automatically performed 3 Display changes to CAL on which shows the mode advanced to span adju...

Страница 23: ...If 1 Err appears on the display the calibration weight is less than 50 of the standard If 2 Err appears on the display an object other than the calibration weight is on the pan or the display tolerance 1 is exceeded Step Action Display 1 Turn on scale 0000 0000 0000 2 Press and hold the F key until the CAL message is displayed CAL 3 Press and hold the T key press the F key then release both keys o...

Страница 24: ...ration weight is less than 50 of the standard If 2 Err appears on the display an object other than the calibration weight is on the pan or the display tolerance 1 is exceeded Step Action Display 1 Turn on scale 00000 00000 00000 2 Press CAL then release the key CAL 3 Press and hold the T key then press the F key then release both keys on 0 4 The scale will automatically be set to the zero point 5 ...

Отзывы: